CALIBRATION SET FOR OEM LV 124 AND OEM LV 148
  • 测试小文本22
  • 测试小文本
  • 应用领域
  • 测试小文本3
  • 符合下列标准
  • 特点
  • 优点
  • 软件
  • 点击下载
Data Sheet CA LV 124 (20140704)_201807121317101.jpg
  • 测试小文本22-1
    测试小文本22
    None
    返回
  • 测试小文本-1
    测试小文本
    None
    返回
  • 应用领域-1
    应用领域

    The CA LV 124 set includes four load resistors for the verification of the pulses E10 and E13 of the LV 124 (2013). The load

    resistors CA LV124-P1R and CA LV124-P100R are used for power line verification. 

    For verification the 16 signal- and datalines the CA LV124-D1R and CA LV124-D1000R are used.


    返回
  • 测试小文本3-1
    测试小文本3
    None
    返回
  • 符合下列标准-1
    符合下列标准

    TEST PROCEDURE

    One reference measurement each with 100 ohm (±5%)

    and 1 ohm (±5%) as a DUT substitute must be performed

    and documented. Verification of the edge steepness must

    be provided with this test setup. Low-inductance parts

    must be used as resistors.

    The load resistor must have the following characteristics

    to achieve the required pulse parameters:

    - Low Inductive Component

    - Shortest possible cables (a few cm) to the PFM output

    - Probe must be as near as possible to the load resistor or,

    as with the CA LV 124-P, using a BNC connection directly

    on the load


    返回
  • 特点-1
    特点

    > Low inductive load resistor for direct connect into

    the PFM 200N100 output power plugs

    > Load resistor 1.0 ohm and 100 ohm for power

    lines verification

    > Load resistor 1.0 ohm and 1000 ohm for datalines

    verification

    > High precision load resistors 1.0, 100 and 1000

    ohms

    > Software procedure for verification with CA LV124

    loads


    返回
  • 优点-1
    优点

    > Low inductive load resistor for direct connect into

    the PFM 200N100 output power plugs

    > Load resistor 1.0 ohm and 100 ohm for power

    lines verification

    > Load resistor 1.0 ohm and 1000 ohm for datalines

    verification

    > High precision load resistors 1.0, 100 and 1000

    ohms

    > Software procedure for verification with CA LV124

    loads


    返回
  • 软件-1
    软件

    POWER LINE SETTINGS

    The reference measurement can be selected by selecting

    Edge Verification in the software. There is a test window

    for call up the reference measurement settings.

    1 Edge Verification: 

    Choose this option to perform the reference

    measurement.

    2 Test Window for the reference measurement


    返回